EasyManua.ls Logo

Fluke 1760 - Page 89

Fluke 1760
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Options and Accessories
Accessories 6
6-11
Crest factor: < 3
Peak current: 18 A, 180 A
Max. non destructive current: Up to 200 A rms
Conductor position influence: < 0.5 % of range at 50/60 Hz
Error due to adjacent conductor: 15 mA/A at 50 Hz
Phase error (to reference conditions): < ±0.5 degrees
Frequency range (clamp without the
instrument):
40 Hz – 10 kHz (3 dB)
Temperature coefficient: 0.015 % of range/ °C
Safety: 300 V AC CAT IV, type C sensor, pollution
degree 2
General Characteristics
Maximum conductor size: Diameter: 15 mm.
Bus bar: 15 x 17 mm
Cable length: 2 m
Operating temperature range: 10 °C - +55 °C
Storage temperature range: 20 - +70 °C
Operating humidity: 15 % - 85 % (non-condensing)
Weight (per clamp): Approx. 220 g
Order-number: 2540461
Reference Conditions
Environment temperature range: +18 °C to +26 °C
Humidity: 20 to 75 % rh
Current: Sinusoidal waveform, with 48 to 65 Hz
Distortion factor: < 1 %, no DC component, stray field < 40 A/m,
conductor centered within the clamp jaws
Safety Standards
IEC/EN 61010-1 2
nd
edition
IEC/EN 61010-2-032
IEC/EN 61010-031
EMC Standards
EN61326 –1:2006
Accuracy (Typical, for 50/60 Hz)
Linearity, error in % of measured value, primary current in A:

Table of Contents

Other manuals for Fluke 1760