EasyManuals Logo

Fluke 177 Calibration Information

Fluke 177
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
True RMS Multimeters
Performance Tests
11
Preparing for the Performance Test
WWarning
To avoid possible electric shock or personal injury:
Do not perform the following procedures unless qualified to do so. Some
procedures involve the use of high voltages.
Before handling the test connections and in between tests, make sure the
calibrator is in standby mode (STBY).
To prepare for the performance test:
1. Make sure that you have the required equipment, see Table 1.
2. Warm up the calibrator as required by its specifications.
3. Allow the temperature of the UUT to stabilize at room temperature ( 23 °C ± 5 °C [73 °F ± 9 °F] ).
4. Check the fuses and battery, and replace them if necessary. Refer to "Replacing the Fuses" and
"Replacing the Battery".
Performance Tests
1. Set the UUT switch to the position called for in Table 2 (175) or Table 3 (177/179), and for all
measurements other than current (amps), connect the calibrator to the VeG and
COM input
terminals.
If testing the milliamps function, connect the calibrator to the meter
400 mA and COM input
terminals.
If testing the amps function, connect the calibrator to the meter
10A and COM input terminals.
2. Referring to Table 2 or Table 3, apply the indicated calibrator output voltages to the UUT for each
test.
3. Verify that the UUT displays a reading that is within the limits shown in the Meter Response column.
Table 2. Performance Tests 175
Meter Response
Test
(Switch Position) 5520 Output Lower Limit Upper Limit
300 mV 45 Hz 296.7 mV AC 303.3 mV AC
5 V 500 Hz 4.947 V AC 5.053 V AC
5 V 1 kHz 4.897 V AC 5.103 V AC
50 V 1 kHz 48.97 V AC 51.03 V AC
50 V 45 Hz 49.47 V AC 50.53 V AC
300 V 45 Hz 296.7 V AC 303.3 V AC
100 V 500 Hz 98.7 V AC 101.3 V AC
100 V 1 kHz 97.7 V AC 102.3 V AC
600 V 500 Hz 593.7 V AC 606.3 V AC
K
AC Volts
1000 V 45 Hz 987 V AC 1013 V AC

Other manuals for Fluke 177

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 177 and is the answer not in the manual?

Fluke 177 Specifications

General IconGeneral
Frequency range0.00001 - 100 kHz
DC current range0.00001 - 10 A
DC voltage range0.0001 - 1000 V
Resistance range0.1 - 50000000 Ω
Resistance resolution0.1 Ω
Resolution (frequency)0.01 Hz
Resolution (AC current)10 µA
Resolution (DC voltage)0.1 mV
Basic accuracy (frequency)±(0.1% + 1)
Basic accuracy (DC voltage)±(0.09% + 2)
Basic accuracy (resistance)±(0.9% + 1)
Basic accuracy (capacitance)±(1.2% + 2)
Battery type9V
Battery voltage9 V
Number of batteries supported1
ManualYes
Weight and Dimensions IconWeight and Dimensions
Depth43 mm
Width90 mm
Height185 mm
Weight420 g

Related product manuals