175, 177, & 179
Calibration Information
4
Accuracy ± ( [ % of Reading ] + [ Counts ] )
Function Range
1
Resolution
Model 175 Model 177 Model 179
AC Volts
2
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
DC mV 600.0 mV 0.1 mV 0.15 % + 2 0.09 % + 2 0.09 % + 2
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.15 % + 2 0.09 % + 2 0.09 % + 2
1000 V 1 V 0.15 % + 2 0.15 % + 2 0.15 % + 2
Continuity 600 1 Meter beeps at < 25 , beeper turns off at > 250 ; detects
opens or shorts of 250 µs or longer.
Ohms 600.0
6.000 k
60.00 k
600.0 k
6.000 M
50.00 M
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance 1000 nF
10.00 µF
100.0 µF
9999 µF
3
1 nF
0.01 µF
0.1 µF
1 µF
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
AC Amps
(True RMS)
(45 Hz to 1 kHz)
60.00 mA
400.0 mA
(600 mA for 18 hrs)
6.000 A
10.00 A
(20 A for 30 s)
0.01 mA
0.1 mA
0.001 A
0.01 A
1.5 % + 3 1.5 % + 3 1.5 % + 3
1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Crest factor of 3 at full scale up to 500 V, decreasing linearly to crest factor 1.5 at 1000 V.
3. In the 9999 µF range for measurements to 1000 µF, the measurement accuracy is 1.2 % for all models.