187/189
Calibration Manual
4
Specifications
Accuracy
Accuracy is specified for a period of one year after calibration, at 18 °C to 28 °C (64 °F
to 82 °F), with relative humidity to 90 %. Accuracy specifications are given as:
± ( [ % of reading ] + [ number of least significant digits ] )
AC mV, AC V, AC µA, AC mA, and AC A specifications are ac-coupled, true rms and are
valid from 5 % of range to 100 % of range. AC crest factor can be up to 3.0 at full-scale, 6.0
at half-scale, except the 3000 mV and 1000 V ranges where it is 1.5 at full scale, 3.0 at half-
scale.
Feature Summary
Feature Description
Dual Digital Displays
Analog Bar Graph
Primary: 50,000 counts
Secondary: 5,000 count
Bar graph: 51 segments, updates 40 times/second
Backlight with 2 brightness levels Bright white backlight for clear readings in poorly lighted areas
Fast Autorange Meter automatically selects best range - instantly
AC+DC true rms, ac rms specified
to 100 kHz
Choices for AC only, AC and DC dual display, or AC+DC
readings
dBm, dBV User selectable impedance references for dBm
AutoHOLD Holds readings on display
Continuity / Open test Beeper sounds for resistance readings below threshold, or to
indicate a momentary open circuit
Fast Bar Graph 51 segments for peaking and nulling
Duty cycle / Pulse width Measure signal on or off time in % or milliseconds
MIN MAX Mode
FAST MN MX with 24-hour time
stamp
Record maximum, minimum, and average values.
24-hour clock for MAX or MIN, elapsed time for AVG.
FAST MN MX captures peaks to 250 µsec.
Closed-Case Calibration No internal adjustments needed
Battery / Fuse Access Door Battery or fuse replaceable without voiding calibration
Hi-Impact Overmolded Case Protective holster features