187/189
Calibration Manual
10
Required Equipment
Required equipment is listed in Table 2. If the recommended models are not available,
equipment with equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table 2. Required Equipment
Equipment Required Characteristics
Recommended
Model
Calibrator AC Voltage Range: 0-1000 Vac
Accuracy: + 0.1 %
Frequency Range: 20 Hz – 100 kHz
Accuracy: +3 %
DC Voltage Range: 0-1000 Vdc
Accuracy: +
0.006 %
Current Range: 0-10A
Accuracy: AC mA - +
0.18 %
AC A - + 0.38 %
Frequency range: 1 kHz – 20 kHz
DC mA - + 0.06 %
DC A - + 0.125 %
Frequency Source: 500 Hz-20 kHz
Accuracy: +
0.00125 %
Amplitude: 1 Volt
Accuracy: + 5 %
Ohms Range: 0 Ω - 50 MΩ
Accuracy: +
0.0125 %
Capacitance Range: 110.0 nF & 11.00
µF
Accuracy: +0.25%
Temperature Range: -10 to 350 °C
Accuracy: 0.3%
Fluke 5520A
Thermocouple Adapter
Accessory
K-type Fluke 80AK
K-type Thermocouple K-type, mini-plug on both ends N/A
Serial Cable, Infrared Fluke P/N 1578406
87/89-IV, 187/189
Service Software
Fluke P/N 676152
IR Cable Adapter Fluke P/N 1576688