187/189
Calibration Manual
18
Testing Voltage, Current, Resistance, Capacitance and Diode Functions
To verify the accuracy of Meter functions, do the following.
1. Connect the Calibrator to the
z and COM inputs on the Meter.
2. Turn the rotary switch to the function listed in each step.
3. Apply the input from the steps listed in Table 4.
4. Compare the reading on the Meter display with the display reading in Table 4.
5. If the display reading falls outside of the range shown in Table 4, the Meter does not
meet specification.
6. Repeat steps 2 through 4 for the remaining inputs shown in Table 4.
Table 4. Performance Test Steps
Display
2
Step Function Range
1
Input Level
Frequency or
Model
Lower
Limit
Upper
Limit
1 AC V 5.0000 V 0.25 V 100 kHz 0.2260 0.2740
2 AC V 5.0000 V 5 V 20 kHz 4.9210 5.0790
3 AC V 5.0000 V 5 V 100 kHz 4.5960 5.4040
4 AC V 50.000 V 50 V 700 Hz 49.760 50.240
5 AC V 50.000 V 50 V 20 kHz 49.210 50.790
6 AC V 50.000 V 50 V 100 kHz 45.960 54.040
7 AC V 500.00 V 500 V 65 Hz 497.60 502.40
8 AC V 1000.0 V 500 V 65 Hz 494.0 506.0
9 AC mV 50.000 2.5 mV 100 kHz 2.085 2.915
10 AC mV 50.000 50 mV 20 kHz 46.850 53.150
11 AC mV 50.000 50 mV 100 kHz 42.460 57.540
12 AC mV 500.00 mV 500 mV 10 kHz 474.60 525.40
13 AC mV 3000.0 mV 3 V 20 kHz 2951.0 3049.0
14 AC mV 3000.0 mV 3 V 100 kHz 2756.0 3244.0
15 Frequency 50.000 kHz 1 V 20 kHz 19.998 20.002
16 Duty Cycle 3.0000 V 2.2 Vp-p,
@30%
Sq. wave
500 Hz
21.82 38.18
17 DC V 5.0000 V short -0.0010 0.0010
18 DC V 5.0000 V 5 V 4.9977 5.0023
19 DC V 5.0000 V -5 V -4.9977 -5.0023
20 DC V 50.000 V -50 V -49.982 -50.018
21 DC V 500.00 V -500 V -499.48 -500.52
22 DC V 1000.0 V -500 V -499.3 -500.7
1. Do not rely on autoranging for all tests. Some test steps will require manual ranging.
2. Must allow for source and meter settling before recording reading.
3. Wait at least 60 seconds for source and meter to settle before recording reading.
4. Using Relative Mode (REL) to zero residual reading (apply 0 and allow UUT to settle for 45 seconds before
pushing REL).