15B+/17B+/18B+
Calibration Manual
10
Disassembly
XWWarning
To prevent false readings, which could lead to possible electric
shock or personal injury, replace the batteries as soon as the
low battery indicator () appears.
To prevent damage or injury:
Install ONLY replacement fuses with the specified
amperage, voltage, and interrupt ratings.
Disconnect test leads before opening the case or the
battery door.
To remove the battery door and battery for adjustments and calibration, see
Figure 1.
icr01.eps
Figure 1. Remove the battery
Performance Tests
Use the performance tests to make sure the UUT is operating properly and to
make sure the UUT is accurate. If the UUT fails any part of the performance test,
repair and/or calibration adjustment are required. See the Calibration sections for
calibration adjustment. If the UUT still fails to meet the range indicated, see How
to Contact Fluke.
Table 2 lists the equipment required for the performance tests.
Table 2. Required Equipment for Performance Test 15B+ and 17B+
Equipment Recommended Model
Calibrator FLUKE 5522A