EasyManua.ls Logo

Fluke 18B+ - Service and Parts Information

Fluke 18B+
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15B+/17B+/18B+
Calibration Manual
16
Connect the UUT signal input terminal to the calibrator (5522A or other
calibrator.)
icr07.eps
Figure 4. Calibration Connections
4. To put the UUT in calibration mode, use a small probe to push the calibration
button (CAL1) See Figure 5 for 15B+ and 17B+ and Figure 7 for 18B+.
Calibration for 15B+ and 17B+
For each function in Table 7:
1. Turn the rotary selection knob to the function to be calibrated.
2. Set calibrator output for the appropriate signal.
3. Wait 4 seconds for the reading to stabilize then push to confirm and
forward next step.
4. On the PCB, short CAL1 and WP6 together. See Figure 5.

Other manuals for Fluke 18B+

Related product manuals