Performance Test for 18B+ 
Before doing the performance test, make sure to warm up the calibrator. 
To test each function and operating ranges: 
1.  For each step in Table 4, set the UUT to the specified function and range.  
2.  Connect the source to the input jacks on the UUT. 
3.  Apply the output from the source. 
4.  Make sure the readings on the UUT display are within the limits in Table 4. 
5.  For the LED test, put the LEDs as shown in Figure 2. The test passes if the 
LED lights. 
Table 4. Performance Specifications for 18B+ 
Step Function 
Calibrator Output  UUT reading limit 
Value 
Frequency or 
Amplitude 
Lower Limit  Upper Limit 
1 
VAC 
3.5 V  50 Hz  3.462  3.538 
2  35 V  50 Hz  34.62  35.38 
3  35 V  500 Hz  34.62  35.38 
4  350 V  50 Hz  346.2  353.8 
5  1000 V  50 Hz  986  1014 
6 
VAC/Freq. 
45 Hz  1 V  44.92  45.08 
7  5 KHz  1 V  4.992  5.008 
8  100 KHz  3 V  99.6  100.4 
9  VAC/Duty  50 Hz 
3 V (Square 
Wave) 
49.5 50.5 
10 
VDC 
3.5 V  0 Hz  3.480  3.520 
11  35 V  0 Hz  34.80  35.20 
12  350 V  0 Hz  348.0  352.0 
13  -1000 V  0 Hz  -1008  -992 
14  mVDC  350 mV  0 Hz  345.4  354.6 
15  mVAC  350 mV  500 Hz  339.2  360.8