Table 4. Performance Specifications for 18B+ (cont.)
Step Function
Calibrator Output UUT reading limit
Value
Frequency or
Amplitude
Lower Limit Upper Limit
16
OHM
0 Ω Ω -0.3 0.3
17 350 Ω Ω 348.0 352.0
18 3.5 kΩ Ω 3.480 3.520
19 35 kΩ Ω 34.80 35.20
20 350 kΩ Ω 348.0 352.0
21 3.5 mΩ Ω 3.480 3.520
22 10 mΩ Ω 9.82 10.18
23 BEEPER ON 40 Ω Ω X X
24 DIODE 0.7 V 0 Hz 0.622 0.778
25
CAP
35 nF Ω 32.75 37.25
26 350 nF Ω 342.4 357.6
27 3.5 μF Ω 3.320 3.680
28 35 μF Ω 33.20 36.80
29 350 μF Ω 332.0 368.0
30
ADC
3.5 A 0 Hz 3.444 3.556
31 -10 A 0 Hz -10.18 -9.82
32
AAC
3.5 A 50 Hz 3.444 3.556
33 10 A 400 Hz 9.82 10.18
34
mADC
35 mA 0 Hz 34.44 35.56
35 -350 mA 0 Hz -355.6 -344.4
36
mAAC
35 mA 400 Hz 34.44 35.56
37 350 mA 50 Hz 344.4 355.6
38
μADC
350 μA 0 Hz 344.4 355.6
39 -3500 μA 0 Hz -3556 -3444
40
μAAC
350 μA 40 Hz 344.4 355.6
41 3500 μA 400 Hz 3444 3556
42
LED
-5.5 V 0 Hz -6.05 -4.95
43 5.5 V 0 Hz 4.95 6.05