Performance Tests
11
Performance Tests for the 15B+ and 17B+
Before doing the performance test, make sure to warm up the calibrator.
To test each function and operating ranges:
1. For each step in Table 3, set the UUT to the specified function and range.
2. Connect the source to the input jacks on the UUT.
3. Apply the output from the source.
4. Make sure the readings on the UUT display are within the limits in Table 3.
Table 3. Performance Specifications for 15B+ and 17B+
Step Function
Calibrator output
15B+ 17B+
UUT reading limit
Value
Frequency
or Amplitude
Lower
Limit
Upper
Limit
1
VAC
3.5 V 50 Hz 3.462 3.538
2 35 V 50 Hz 34.62 35.38
3 35 V 500 Hz 34.62 35.38
4 350 V 50 Hz 346.2 353.8
5 1000 V 50 Hz 986 1014
6
VAC/Freq.
45 Hz 1 V X 44.92 45.08
7 5 KHz 1 V X 4.992 5.008
8 100 KHz 3 V X 99.6 100.4
9 VAC/Duty 50 Hz
3 V (Square
Wave)
X 49.5 50.5
10
VDC
3.5 V 0 Hz 3.480 3.520
11 35 V 0 Hz 34.80 35.20
12 350 V 0 Hz 348.0 352.0
13 -1000 V 0 Hz -1008 -992
14 mVDC 350 mV 0 Hz 345.4 354.6
15 mVAC 350 mV 500 Hz 339.2 360.8