EasyManua.ls Logo

Fluke 18B+ - Measure Frequency and Duty Cycle (17 B+;18 B+ Only); Test LEDs (18 B+ Only)

Fluke 18B+
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15B+/17B+/18B+
Calibration Manual
12
Table 0-3. Performance Specifications for 15B+ and 17B+ (cont.)
Step Function
Calibrator output
15B+ 17B+
UUT reading limit
Value
Frequency
or Amplitude
Lower
Limit
Upper
Limit
16
OHM
0 Ω Ω -0.3 0.3
17 350 Ω Ω 348.0 352.0
18 3.5 kΩ Ω 3.480 3.520
19 35 kΩ Ω 34.80 35.20
20 350 kΩ Ω 348.0 352.0
21 3.5 MΩ Ω 3.480 3.520
22 10 MΩ Ω 9.82 10.18
23 BEEPER ON 40 Ω Ω X X
24 DIODE 0.7 V 0 Hz 0.622 0.778
25
CAP
35 nF Ω 34.24 35.76
26 350 nF Ω 342.4 357.6
27 3.5 μF Ω 3.320 3.680
28 35 μF Ω 33.20 36.80
29 350 μF Ω 332.0 368.0
30
ADC
3.5 A 0 Hz 3.444 3.556
31 -10 A 0 Hz -10.18 -9.82
32
AAC
3.5 A 50 Hz 3.444 3.556
33 10 A 400 Hz 9.82 10.18
34
mADC
35 mA 0 Hz 34.44 35.56
35 -350 mA 0 Hz -355.6 -344.4
36
mAAC
35 mA 400 Hz 34.44 35.56
37 350 mA 50 Hz 344.4 355.6
38
μADC
350 μA 0 Hz 344.4 355.6
39 -3500 μA 0 Hz -3556 -3444
40
μAAC
350 μA 40 Hz 344.4 355.6
41 3500 μA 400 Hz 3444 3556
42
Temp
-50 °C X X -56.4 -43.6
43 35 °C X X 33.0 37.0
44 400 °C X X 391.0 409.0

Other manuals for Fluke 18B+