Digital Multimeter
Calibration Procedures
15
icr08.eps
Figure 2. LED Test
Calibration Procedures
Table 5 list equipment that is required to calibrate the Product.
Table 5. Required Equipment for Calibration
Equipment Required Characteristics Recommended Model
Calibrator FLUKE 5522A
Potentiometer (17B+) Manually adjustable
To enter calibration mode:
1. Remove the battery door and battery. See Figure 1.
2. Remove the calibration sticker. See Figure 3.
icr06.eps
Figure 3. Remove Calibration Sticker
3. See Figure 4 to make the connections to power and a calibrator:
• Use a contact spring to connect to a 3 V dc power supply.