233
Calibration Information
14
Table 3. Performance Tests (cont)
Step Function Range Applied Display Reading
19 6.000V 0V
-0.002 to 0.002
20 6.000V 5V 4.985 to 5.015
21 60.00V 50V 49.85 to 50.15
22 600.0V 600V 598.3 to 601.7
23 600.0V -600V -598.3 to -601.7
24
L
DC Volts
1000V 1000V 995 to 1000
25 600.0 mV 6 mV, 45 Hz 5.6 to 6.4
26
m
AC Millivolts
600.0 mV 600mV, 1 kHz 587.7 to 612.3
27 600.0 mV 20 mV 19.7 to 20.3
28
E
DC Millivolts
600.0 mV 600mV 598.3 to 601.7
29 G Diode 2.000 V 1.9V
1.881 to 1.919
31 1000 nF 30
27 to 33
32 9 μF 8.81 to 9.19
35 90 μF 88.1 to 91.9
36
E
Capacitance
9999 μF 900 μF 881 to 919
Set calibrator to standby, reconfigure leads, and program for amps output
37 A
DC Amps
10.00 A 9.5 A 9.37 to 9.63
38 ?
AC Amps
6.000 A 5.0 A, 45 Hz 4.922 to 5.078
39 Open input 0PEn
40 0.0 °C -1.0 to 1.0
41
y
Temperature
400 °C 395.0 to 405.0
[1] If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency.
Calibration Adjustment
The Meter features closed-case calibration adjustment using known reference sources. The Meter measures the
applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory.
The following sections present the features and Meter pushbutton functions available during the Calibration
Adjustment Procedure. Should the Meter fail any of the performance tests, perform the Calibration Adjustment
Procedure.
Use the following steps to view the Meter’s calibration counter.
1. While pressing K, turn the rotary switch from OFF to Ω function. The Meter should display
CAL”.
2. Press g once to view the calibration counter. For example, “n001