233
Calibration Information
18
Table 4. Calibration Adjustment Steps
Rotary
Switch
Position
Calibration Steps
Input
Terminals
Calibrator
Source Value
C-01 VΩ/+ and COM 0 V, 0 Hz
C-02 VΩ/+ and COM 300 mV, 0 Hz
C-03 VΩ/+ and COM 100 mV, 0 Hz
C-04 VΩ/+ and COM -300 mV, 0 Hz
C-05 VΩ/+ and COM 60 mV, 0 Hz
C-06 VΩ/+ and COM 600 mV, 0 Hz
m
C-07 VΩ/+ and COM 600 mV, 60 Hz
C-08 VΩ/+ and COM 600 e, 2-wire comp
C-09 VΩ/+ and COM 6 ke
C-10 VΩ/+ and COM 60 ke
C-11 VΩ/+ and COM 600 ke
C-12 VΩ/+ and COM 6 Me
[1]
C-13 VΩ/+ and COM Short
[1]
e
Ohms
C-14 VΩ/+ and COM 40 Me
[1]
C-15 VΩ/+ and COM 6 V, 60 Hz
C-16 VΩ/+ and COM 60 V, 60 Hz
C-17 VΩ/+ and COM 600 V, 60 Hz
C-18 VΩ/+ and COM 6 V, 0 Hz
C-19 VΩ/+ and COM 60 V, 0 Hz
K
C-20 VΩ/+ and COM 600 V, 0 Hz
Set calibrator to standby, reconfigure leads, and program for amps output.
? C-22 A and COM 6 A, 60 Hz
[1]
A C-23 A and COM 6 A, 0 Hz
[1] Wait an additional 5 seconds after calibrator has settled before pressing g.