Digital Multimeters
11
1
4
5
6
2
3
gaq10.eps
Figure 2. Battery and Fuse Replacement
Performance Tests
XWWarning
To avoid electric shock, do not perform the performance test
procedures unless the Meter is fully assembled.
The following performance tests verify the complete operation of the Meter and check
the accuracy of each Meter function against its specifications. Performance tests should
be performed bi-annually to ensure that the Meter is within accuracy specifications. If the
Meter fails any part of the test, calibration adjustment and/or repair is indicated.
In the performance tests, the Meter is referred to as the unit under test (UUT).