6. Verify that the UUT indicates between 196.8 and 203.2 mA.
7. Program the DMM Calibrator output for zero amps.
8. Switch the UUT to the ac µO function (µO), then program the DMM Calibrator
output for 2000 µO at 500 Hz.
9. Verify that the UUT indicates between 1968 and 2032 µA.
10. Program the DMM Calibrator to standby.
11. Connect the DMM Calibrator output to the UUT amp (A) and common (COM) input
terminals.
12. Switch the UUT to the AC mA/A function, then program the DMM Calibrator
output for 5 A at 1 kHz.
13. Verify that the UUT indicates between 4.91 and 5.09 A. Program the DMM
Calibrator to standby.
14. Set the UUT to the dc mN/A function(mN/A).
15. Program the DMM Calibrator output for an output of 5.0 A dc.
16. Verify that the UUT indicates between 4.94 and 5.06 A.
17. Program the DMM Calibrator to standby.
18. Turn the function switch on the UUT to the dc mN/A position (mN/A).
19. Connect the DMM Calibrator output to the UUT mAA input terminal, then
program the DMM Calibrator for an output of 27 mA dc.
20. Verify that the UUT indicates between 26.78 and 27.22 mA.
21. Program the DMM Calibrator output for 2700 µN dc. Set the UUT to the dc µN
function (µN).
22. Program the DMM Calibrator for 2.7 mA.
23. Verify that the UUT indicates between 2678 and 2722 µA.
3-16. Ohms Function Performance Verification
The following test may be used to verify correct ohms function operation and to verify
UUT accuracy in the various ohms ranges.
1. Connect the DMM Calibrator to the UUT using the volts/ohms/diode-test input
terminal and common.
2. Turn the function switch to the J position.
3. Switch on power to the DMM Calibrator. Program the DMM Calibrator to the
resistance values indicated in Table 3-2, and verify that the UUT indication is within
the tolerances given for each input value.

Other manuals for Fluke 27

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 27 and is the answer not in the manual?

Fluke 27 Specifications

General IconGeneral
Display TypeLCD
AC Voltage Range0.1 mV to 1000 V
DC Voltage Range0.1 mV to 1000 V
AC Current Range0.1 μA to 10 A
DC Current Range0.1 μA to 10 A
Battery Type9V
ContinuityYes
Diode TestYes
Safety RatingCAT III 600V
AC Current Accuracy±(1.5% + 2 digits)
Resistance Accuracy±(0.2% + 1 digit)
DC Voltage Accuracy±(0.1% + 1 digit)

Related product manuals