EasyManuals Logo

Fluke 45 User Manual

Fluke 45
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
Operating the Meter From the Front Panel
Power-Up Configuration
3
3-19
The last number entered (or, in REL, the last measured value) is shown with the left most
digit displayed in normal intensity and the remaining digits dim. (If the number is
negative, the minus sign is dim.) Related annunciators are lit, and "EDIT" is shown in the
secondary display.
2. Press U to step the bright digit through numbers between O and 9. Holding down U
causes this digit to scroll continuously.
Press T to select the digit to be edited. Each press of T causes the digit to the right
to be selected and go bright, while causing the previously selected digit to go dim. When
the right most digit has been selected, the next press of T selects the sign: if the sign is
positive, the negative annunciator flashes bright dim; if the sign is negative, the negative
annunciator is bright. U will toggle the sign between positive and negative.
3. Any time during editing, you can store a selected value by pressing E.
4. To abort the edit and return the meter to normal operation, press any button except E,
U, and T .
Power-Up Configuration
Factory Settings of Power-Up Configuration
When the meter is turned on and completes the power-up sequence, it assumes its power-
up configuration. The power-up configuration set at the factory is shown in Table 3-13.
The IEEE-488 address, the RS-232 baud rate, parity, echo, and print-only rate are not
changed when power is cycled off and on. These parameters remain as set until changed
by the user.
Changing the Power-Up Configuration
You can change the power-up configuration to one that more closely meets your needs
and preferences. Any combination of meter parameters can become the power-up
configuration.
To save any configuration in which the meter is operating, press S and, while holding
it down, press E. The following parameters are saved:
Measurement function and initial range on primary display.
Measurement function and initial range on secondary display.

Table of Contents

Other manuals for Fluke 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 45 and is the answer not in the manual?

Fluke 45 Specifications

General IconGeneral
Diode TestYes
Display Digits5 digits
InterfaceRS-232
Frequency Range5Hz to 500kHz
Continuity< 25 Ω
Basic DC Accuracy0.05%
Power Supply115/230 V AC, 50/60 Hz

Related product manuals