EasyManua.ls Logo

Fluke 45 - 4-6. Entering Calibration Mode

Fluke 45
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating the Meter Using the Computer Interface
Computer Interface Command Set
5
5-19
6 will always be zero.) Convert the returned value to binary to determine the status of
register bits.
EXAMPLE EXPLANATION
*SRE? Reads the value of the SRE Register. Assume "32" is returned.
Converting 32 to the binary 00100000 indicates that bit 5 in the SRE
is set to 1.
Computer Interface Command Set
The remainder of Chapter 5 describes the RS-232 and IEEE-488 computer interface
commands. RS-232 and IEEE-488 commands are identical, except where indicated.
These commands, grouped by related function, are listed in the tables that follow. A
parameter that must be supplied by the user, or a string returned by the meter is enclosed
in angle brackets (e.g., <value>).
IEEE-488 Capabilities and Common Commands (Tables 5-8 and 5-9)
Function Commands and Queries (Table 5-11)
Function Modifier Commands and Queries (Table 5-12)
Range and Measurement Rate Commands and Queries (Table 5-13)
Measurement Queries (Table 5-14)
Compare Commands and Queries (Table 5-15)
Trigger Configuration Commands (Table 5-16)
Miscellaneous Commands and Queries (Table 5-17)
RS-232 Remote/ Local Configuration Commands (Table 5-18)
IEEE-488 Capabilities and Common Commands
The meter supports the IEEE-488 capabilities shown in Table 5-8.
Table 5-9 Summarizes the IEEE-488 Common Commands.

Table of Contents

Other manuals for Fluke 45

Related product manuals