EasyManua.ls Logo

Fluke 540B - Page 74

Fluke 540B
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHANGE
#1
-
3071
On page 5-14, delete the entries for the following items:
Shunt Retaining Screw
Holder
Knob
On page 5-15, delete the following
part
numbers and corresponding mechanical parts from Figure 5-6:
2405-158949
2102-103283
3153-177485
CHANGE #2 -
8021
On
page 5-12, delete the existing entries for diodes CR307 and CR314 and replace each with
th
e following:
Diode, Si; 4805-375907; 07263; FD7222; (CR307)
2,
(CR314)
REF
CHANGE #3 - 11053
On page 5-10, change the description, Fluke stock
no
, mfr,
and
mfr part
no
for transistors Q7 and
Q9
From
: type 2N1307,
4805-148643,01295,
2Nl307
To: Si, PNP,
4818
-341081; 07263; 2N4121
ERRATA
#1
540B
The
540B-110
Rechargeable Battery Pack
is
described
in
the manual as
an
accessory. Change the manual
to
indicate
that the standard
540B
now
includes the Rechargeable Battery Pack.
ERRATA
#2
On page 5-
14
and 5-15 change the Fluke stock
no
and
mfr part
no
for the Range Switch Cover from 3206-178459
to
315
5-1
784
67.
ERRATA
#3
On
page 2-5, paragraph 2-10-g, change the last sentence to read:
....
. . and the device should be rechecked for
heater-to-couple isolation (see Section
4).
Defective elements should be replaced.
On page 4-5, paragraph 4-23, chan
ge
the last sentence
to
read:
.....
.
to
replace the
th
erm
ocouple. Replacement
of
the thermocouple should
not
require recalibration beyond the DC Rerversal and Heater-to-Couple Isolation tests.
ERRATA
#4
On
page 5-
3,
change the Fluke stock
no
and mfr
part
no
for the Magnetic Modulator/Carrier Amplifier Assembly
from 1702-165803
to
1702-242495. Also change
840A401
in
the Fluke stock
no
column to
540B-410
.
On page 5-l b correct the Magnetic Modulator/Carrier Amplifier Assembly's
mfr
part
no
from 1 702-165803
to
1702·
242495.
11n1