Manual Supplement 5520A Operator
7/11 1
Change #1,
29572,31572,39294,46109,49164,51096,51915,57255,57658,57928,58756,58937,58991,59053
On pages 1-9 through 1-34, replace the entire Specifications with the following:
Specifications
The following tables list the 5520A specifications. All specifications are valid after allowing a warm-up period of 30 minutes, or twice the time
the 5520A has been turned off. (For example, if the 5520A has been turned off for 5 minutes, the warm-up period is 10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures outside of t
cal
±5 °C (t
cal
is the ambient temperature
when the 5520A was calibrated), the temperature coefficient as stated in the General Specifications must be applied.
The specifications also assume the Calibrator is zeroed every seven days or whenever the ambient temperature changes more than 5 °C. The
tightest ohms specifications are maintained with a zero cal every 12 hours within ±1 °C of use.
Also see additional specifications later in this chapter for information on extended specifications for ac voltage and current. The dimensional
outline for the 5520A Calibrator is shown in Figure A.
POWER
I
O
0•
123
456
789
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH EXGRD SCOPE MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
μ
n
p
W
dBm sec
¡CA
MULT
x
DIV
÷
MODES
MORE
STBY
43.2 cm (17 in)
47.0 cm (18.5 in) 6.4 cm
(2.5 in)
For Cable
Access
17.8 cm
(7 in)
HI
LO
TRIG
GUARD
TC
20A
NORMAL AUX
5520A CALIBRATOR
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
20V PK MAX
20V PK MAX
nn032f.eps
Figure A. 5520A Calibrator Dimensional Outline