EasyManua.ls Logo

Fluke 5520A - AC Voltage (Sine Wave) Specifications

Fluke 5520A
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Supplement 5520A Operator
7/11 7
AC Voltage (Sine Wave) Specifications
NORMAL (Normal Output)
Range Frequency
Absolute Uncertainty,
tcal ±5 °C
±(ppm of output + μV)
Resolution
Max
Burden
Max Distortion and
Noise
10 Hz to 5 MHz
Bandwidth
±(% of output + floor)
90 days 1 year
1.0 to
32.999 mV
10 to 45 Hz
600 + 6 800 + 6
1 μV 50 Ω
0.15 + 90 μV
45 Hz to 10 kHz
120 + 6 150 + 6 0.035 + 90 μV
10 to 20 kHz
160 + 6 200 + 6 0.06 + 90 μV
20 to 50 kHz
800 + 6 1000 + 6 0.15 + 90 μV
50 to 100 kHz
3000 + 12 3500 + 12 0.25 + 90 μV
100 to 500 kHz
6000 + 50 8000 + 50 0.3 + 90 μV
[1]
33 to
329.999 mV
10 to 45 Hz
250 + 8 300 + 8
1 μV 50 Ω
0.15 + 90 μV
45 Hz to 10 kHz
140 + 8 145 + 8 0.035 + 90 μV
10 to 20 kHz
150 + 8 160 + 8 0.06 + 90 μV
20 to 50 kHz
300 + 8 350 + 8 0.15 + 90 μV
50 to 100 kHz
600 + 32 800 + 32 0.20 + 90 μV
100 to 500 kHz
1600 + 70 2000 + 70 0.20 + 90 μV
[1]
0.33 to
3.29999 V
10 to 45 Hz
250 + 50 300 + 50
10 μV
10 mA
0.15 + 200 μV
45 Hz to 10 kHz
140 + 60 150 + 60 0.035 + 200 μV
10 to 20 kHz
160 + 60 190 + 60 0.06 + 200 μV
20 to 50 kHz
250 + 50 300 + 50 0.15 + 200 μV
50 to 100 kHz
550 + 125 700 + 125 0.20 + 200 μV
100 to 500 kHz
2000 + 600 2400 + 600 0.20 + 200 μV
[1]
3.3 to
32.9999 V
10 to 45 Hz
250 + 650 300 + 650
100 μV
10 mA
0.15 + 2 mV
45 Hz to 10 kHz
125 + 600 150 + 600 0.035 + 2 mV
10 to 20 kHz
220 + 600 240 + 600 0.08 + 2 mV
20 to 50 kHz
300 + 600 350 + 600 0.2 + 2 mV
50 to 100 kHz
750 + 1600 900 + 1600 0.5 + 2 mV
33 to
329.999 V
45 Hz to 1 kHz
150 + 2000 190 + 2000
1 mV
5
mA,except
20 mA for
45 to 65 Hz
0.15 + 10 mV
1 to 10 kHz
160 + 6000 200 + 6000 0.05 + 10 mV
10 to 20 kHz
220 + 6000 250 + 6000 0.6 + 10 mV
20 to 50 kHz
240 + 6000 300 + 6000 0.8 + 10 mV
50 to 100 kHz
1600 + 50000 2000 + 50000 1.0 + 10 mV
330 to
1020 V
45 Hz to 1 kHz
250 + 10000 300 + 10000
10 mV
2 mA,
except
6 mA for 45
to 65 Hz
0.15 + 30 mV
1 to 5 kHz
200 + 10000 250 + 10000 0.07 + 30 mV
5 to 10 kHz
250 + 10000 300 + 10000 0.07 + 30 mV
[1] Max Distortion for 100 to 200 kHz. For 200 to 500 kHz, the maximum distortion is 0.9 % of output + floor as shown.
Note
Remote sensing is
not provided. Output resistance is  5 mΩ for outputs 0.33 V. The maximum load capacitance is 500 pF, subject to
the maximum burden current limits.

Other manuals for Fluke 5520A