EasyManua.ls Logo

Fluke 5520A - Page 9

Fluke 5520A
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Supplement 5520A Operator
8 7/11
AC Voltage (Sine Wave) Specifications (cont.)
AUX (Auxiliary Output) [dual output mode only]
[1]
Range Frequency
Absolute Uncertainty,
tcal ±5 °C
±(% of output + μV)
Resolution
Max
Burden
Max Distortion and
Noise
10 Hz to 100 kHz
Bandwidth
±
(% of output + floor)
90 days 1 year
10 to
329.999 mV
10 to 20 Hz
0.15 + 370 0.2 + 370
1 μV
5 mA
0.2 + 200 μV
20 to 45 Hz
0.08 + 370 0.1 + 370 0.06 + 200 μV
45 Hz to 1 kHz
0.08 + 370 0.1 + 370 0.08 + 200 μV
1 to 5 kHz
0.15 + 450 0.2 + 450 0.3 + 200 μV
5 to 10 kHz
0.3 + 450 0.4 + 450 0.6 + 200 μV
10 Hz to 30 kHz 4.0 + 900
5.0 + 900 1 + 200 μV
0.33 to
3.29999 V
10 to 20 Hz
0.15 + 450 0.2 + 450
10 μV
5 mA
0.2 + 200 μV
20 to 45 Hz
0.08 + 450 0.1 + 450 0.06 + 200 μV
45 Hz to 1 kHz
0.07 + 450 0.09 + 450 0.08 + 200 μV
1 to 5 kHz
0.15 + 1400 0.2 + 1400 0.3 + 200 μV
5 to 10 kHz
0.3 + 1400 0.4 + 1400 0.6 + 200 μV
10 to 30 kHz
4.0 + 2800 5.0 + 2800 1 + 200 μV
3.3 to 5 V
10 to 20 Hz
0.15 + 450 0.2 + 450
100 μV
5 mA
0.2 + 200 μV
20 to 45 Hz
0.08 + 450 0.1 + 450 0.06 + 200 μV
45 Hz to 1 kHz
0.07 + 450 0.09 + 450 0.08 + 200 μV
1 to 5 kHz
0.15 + 1400 0.2 + 1400 0.3 + 200 μV
5 to 10 kHz
0.3 +1400 0.4 + 1400 0.6 + 200 μV
[1] There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. The AUX output resistance is < 1 Ω. The maximum load capacitance is 500 pF, subject to the
maximum burden current limits.

Other manuals for Fluke 5520A

Related product manuals