Manual Supplement 5520A Operator
18 7/11
AC Voltage (Sine Wave) Extended Bandwidth Specifications
Range Frequency
1-Year Absolute Uncertainty
tcal ±5 °C
Max Voltage
Resolution
Normal Channel (Single Output Mode)
1.0 to 33 mV
0.01 to 9.99 Hz
±(5.0 % of output +
0.5 % of range)
Two digits, e.g., 25 mV
34 to 330 mV Three digits
0.4 to 33 V Two digits
0.3 to 3.3 V
500.1 kHz to 1 MHz -10 dB at 1 MHz, typical
Two digits
1.001 to 2 MHz -31 dB at 2 MHz, typical
Auxiliary Output (Dual Output Mode)
10 to 330 mV
0.01 to 9.99 Hz
±(5.0 % of output +
0.5 % of range)
Three digits
0.4 to 5 V Two digits
AC Voltage (Non-Sine Wave) Specifications
Triangle Wave &
Truncated Sine
Range, p-p
[1]
Frequency
1-Year Absolute Uncertainty,
tcal ±5 °C,
±(% of output + % of range)
[2]
Max Voltage
Resolution
Normal Channel (Single Output Mode)
2.9 to 92.999 mV
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
[3]
5.0 + 0.5
93 to 929.999 mV
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
[3]
5.0 + 0.5
0.93 to 9.29999 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
[3]
5.0 + 0.5
9.3 to 93 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
[3]
5.0 + 0.5
Auxiliary Output (Dual Output Mode)
29 to 929.999 mV
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
5.0 + 0.5
0.93 to 9.29999 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
5.0 + 0.5
9.3 to 14.0000 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
5.0 + 0.5
[1] To convert p-p to rms for triangle wave, multiply the p-p value by 0.2886751. To convert p-p to rms for truncated sine wave,
multiply the p-p value by 0.2165063.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Uncertainty for Truncated Sine outputs is typical over this frequency band.