Manual Supplement 5520A Operator
7/11 19
AC Voltage (Non-Sine Wave) Specifications (cont.)
Square Wave
Range
(p-p)
[1]
Frequency
1-Year Absolute Uncertainty,
tcal ±5 °C
±(% of output + % of range)
[2]
Max Voltage
Resolution
Normal Channel (Single Output Mode)
2.9 to 65.999 mV
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
5.0 + 0.5
66 to 659.999 mV
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
5.0 + 0.5
0.66 to 6.59999 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
5.0 + 0.5
6.6 to 66.0000 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 20 kHz
0.5 + 0.25
20 to 100 kHz
5.0 + 0.5
Auxiliary Output (Dual Output Mode)
29 to 659.999 mV
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
[3]
5.0 + 0.5
0.66 to 6.59999 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
[3]
5.0 + 0.5
6.6 to 14.0000 V
0.01 to 10 Hz
5.0 + 0.5
Two digits on each range
10 to 45 Hz
0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
[3]
5.0 + 0.5
[1] To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2
] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.