7/11 21
AC Current (Non-Sine Wave) Specifications
Triangle Wave &
Truncated Sine Wave
Range
p-p
Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
0.047 to
0.92999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
0.93 to
9.29999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
9.3 to
92.9999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
93 to
929.999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.5
1 to 10 kHz
10 + 2
0.93 to
8.49999 A
[2]
10 to 45 Hz
0.5 + 1.0
Six digits
45 Hz to 1 kHz
0.5 + 0.5
1 to 10 kHz
10 + 2
8.5 to 57 A
[2]
45 to 500 Hz
0.5 + 0.5
500 Hz to 1 kHz
1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
Square Wave
Range p-p
Frequency
1-Year Absolute Uncertainty,
tcal ±5 °C,
±(% of output + % of range)
Max Current
Resolution
0.047 to
0.65999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
0.66 to
6.59999 mA
[1]
10 to 45 Hz
0.25 +0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
6.6 to
65.9999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
66 to
659.999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.5
1 to 10 kHz
10 + 2
0.66 to
5.99999 A
[2]
10 to 45 Hz
0.5 + 1.0
45 Hz to 1 kHz
0.5 + 0.5
1 to 10 kHz
10 + 2
6 to 41 A
[2]
45 to 500 Hz
0.5 + 0.5
500 Hz to 1 kHz
1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.

Other manuals for Fluke 5520A