Manual Supplement 5520A Operator
11/10 25
Trigger Signal Specifications (Edge Function)
Edge Signal
Frequency
Division Ratio
Typical Amplitude into
50 Ω (p-p)
Typical Rise Time Typical Lead Time
900 Hz to 1.1 MHz off/1
1 V 2 ns
40 ns
On page 8-66, in graphic gl021i.eps:
Change:
DC < - > AC
To:
DC - - > AC
On page 8-67, in graphic’s gl023i.eps and gl003i.eps:
Change:
120.00
To:
120.0
On page 8-68, in graphic’s gl004i.eps and gl005i.eps:
Change:
120.00
To:
120.0
On page 8-69, in graphic gl024i.eps:
Change:
DC < - > AC
To:
DC - - > AC
Under Each menu item is described below:
Change:
DC < - > AC
To:
DC < - AC or DC - > AC
On page 8-70, in graphic gl025i.eps:
Change:
DC < - > AC
To:
DC - - > AC
In graphic gl025i.eps, under 20.00 mV/div up/down add 200 to the first column.
On page 8-71, in graphic gl026i.eps:
Change:
DC < - > AC
To:
DC - - > AC
Replace step 2 with the following:
2. Key in the voltage level that is recommended for your oscilloscope. For example to enter 30 mV, press
30gV, then press E. See “Keying in a Value” earlier in this chapter.
Step 3, replace the second paragraph with the following:
This example shows the gain at 30 mV to be 6 divisions, at 5 mV per division.