AC Current (Sine Wave) Specifications
LCOMP off
Range Frequency
Absolute Uncertainty,
tcal ±5 °C
±(% of output + μA)
Compliance
adder
±(μA/V)
Max Distortion &
Noise 10 Hz to
100 kHz BW
±(% of output
+ floor)
Max
Inductive
Load μH
90 days 1 year
29.00 to
329.99 μA
10 to 20 Hz
0.16 + 0.1 0.2 + 0.1
0.05
0.15 + 0.5 μA
200
20 to 45 Hz
0.12 + 0.1 0.15 + 0.1
0.05
0.1 + 0.5 μA
45 Hz to 1 kHz
0.1 + 0.1 0.125 + 0.1
0.05
0.05 + 0.5 μA
1 to 5 kHz
0.25 + 0.15 0.3 + 0.15
1.5
0.5 + 0.5 μA
5 to 10 kHz
0.6 + 0.2 0.8 + 0.2
1.5
1.0 + 0.5 μA
10 to 30 kHz
1.2 + 0.4 1.6 + 0.4
10
1.2 + 0.5 μA
0.33 to
3.29999 mA
10 to 20 Hz
0.16 + 0.15 0.2 + 0.15
0.05
0.15 + 1.5 μA
200
20 to 45 Hz
0.1 + 0.15 0.125 + 0.15
0.05
0.06 + 1.5 μA
45 Hz to 1 kHz
0.08 + 0.15 0.1+ 0.15
0.05
0.02 + 1.5 μA
1 to 5 kHz
0.16 + 0.2 0.2 + 0.2
1.5
0.5 + 1.5 μA
5 to 10 kHz
0.4 + 0.3 0.5 + 0.3
1.5
1.0 + 1.5 μA
10 to 30 kHz
0.8 + 0.6 1.0 + 0.6
10
1.2 + 0.5 μA
3.3 to
32.9999 mA
10 to 20 Hz
0.15 + 2 0.18 + 2
0.05
0.15 + 5 μA
50
20 to 45 Hz
0.075 + 2 0.09 + 2
0.05
0.05 + 5 μA
45 Hz to 1 kHz
0.035 + 2 0.04 + 2
0.05
0.07 + 5 μA
1 to 5 kHz
0.065 + 2 0.08 + 2
1.5
0.3 + 5 μA
5 to 10 kHz
0.16 + 3 0.2 + 3
1.5
0.7 + 5 μA
10 to 30 kHz
0.32 + 4 0.4 + 4
10
1.0 + 0.5 μA
33 to
329.999 mA
10 to 20 Hz
0.15 +20 0.18 + 20
0.05
0.15 + 50 μA
50
20 to 45 Hz
0.075 + 20 0.09 + 20
0.05
0.05 + 50 μA
45 Hz to 1 kHz
0.035 + 20 0.04 + 20
0.05
0.02 + 50 μA
1 to 5 kHz
0.08 + 50 0.10 + 50
1.5
0.03 + 50 μA
5 to 10 kHz
0.16 + 100 0.2 + 100
1.5
0.1 + 50 μA
10 to 30 kHz
0.32 + 200 0.4 + 200
10
0.6 + 50 μA
0.33 to
1.09999 A
10 to 45 Hz
0.15 + 100 0.18 + 100
0.2 + 500 μA
2.5
45 Hz to 1 kHz
0.036 + 100 0.05 + 100
0.07 + 500 μA
1 to 5 kHz
0.5 + 1000 0.6 + 1000
[2]
1 + 500 μA
5 to 10 kHz
2.0 + 5000 2.5 + 5000
[3]
2 + 500 μA
1.1 to
2.99999 A
10 to 45 Hz
0.15 + 100 0.18 + 100
0.2 + 500 μA
2.5
45 Hz to 1 kHz
0.05 + 100 0.06 + 100
0.07 + 500 μA
1 to 5 kHz
0.5 + 1000 0.6 + 1000
[2]
1 + 500 μA
5 to 10 kHz
2.0 + 5000 2.5 + 5000
[3]
2 + 500 μA
3 to
10.9999 A
45 to 100 Hz
0.05 + 2000 0.06 + 2000
0.2 + 3 mA
1
100 Hz to 1 kHz
0.08 + 2000 0.10 + 2000
0.1 + 3 mA
1 to 5 kHz
2.5 + 2000 3.0 + 2000
0.8 + 3 mA
11 to
20.5 A
[1]
45 to 100 Hz
0.1 + 5000 0.12 + 5000
0.2 + 3 mA
1
100 Hz to 1 kHz
0.13 + 5000 0.15 + 5000
0.1 + 3 mA
1 to 5 kHz
2.5 + 5000 3.0 + 5000
0.8 + 3 mA
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure B. The current may be provided 60-T-I
minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in
Amps. For example, 17 A, at 23 °C could be provided for 60-23-17 = 20 minutes each hour. When the 5520A is outputting
currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the
allowable "on" time indicated by the formula and Figure B is achieved only after the 5520A is outputting currents <5 A for the "off"
period first.
[2] For compliance voltages greater than 1 V, add 1 mA/V to the floor specification from 1 to 5 kHz.
[3] For compliance voltages greater than 1 V, add 5 mA/V to the floor specification from 5 to 10 kHz.