EasyManua.ls Logo

Fluke 5520A - Electrical Specifications

Fluke 5520A
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Supplement 5520A Operator
7/11 3
Electrical Specifications
DC Voltage Specifications
Range
Absolute Uncertainty, tcal
±5 °C
±(ppm of output + μV)
Stability
Resolution
μV
Max
Burden
[1]
90 days 1 year
24 hours, ±1 °C
±(ppm output + μV)
0 to 329.9999 mV
15 + 1 20 + 1 3 + 1
0.1
50 Ω
0 to 3.299999 V
9 + 2 11 + 2 2 + 1.5
1 10 mA
0 to 32.99999 V
10 + 20 12 + 20 2 + 15
10 10 mA
30 to 329.9999 V
15 + 150 18 + 150 2.5 + 100
100 5 mA
100 to 1000.000 V
15 + 1500 18 + 1500 3 + 300
1000 5 mA
Auxiliary Output (dual output mode only)
[2]
0 to 329.999 mV
300 + 350 400 + 350 30 + 100
1 5 mA
0.33 to 3.29999 V
300 + 350 400 + 350 30 + 100
10 5 mA
3.3 to 7 V
300 + 350 400 + 350 30 + 100
100 5 mA
TC Simulate and Measure in Linear 10 μV/°C and 1 mV/°C modes
[3]
0 to 329.999 mV
40 + 3 50 + 3 5 + 2
0.1
10 Ω
[1] Remote sensing is not provided. Output resistance is < 5 mΩ for outputs 0.33 V. The AUX output has an output resistance of
<1 Ω. TC simulation has an output impedance of 10 Ω ± 1 Ω.
[2] Two channels of dc voltage output are provided.
[3] TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 V/m.
Range
Noise
Bandwidth 0.1 Hz to
10 Hz p-p
±(ppm of output + floor)
Bandwidth 10 Hz to 10 kHz rms
0 to 329.9999 mV
0 + 1 μV 6 μV
0 to 3.299999 V
0 + 10 μV 60 μV
0 to 32.99999 V
0 + 100 μV 600 μV
30 to 329.9999 V
10 + 1 mV
20 mV
100 to 1000.000 V
10 + 5 mV
20 mV
Auxiliary Output (dual output mode only)
[1]
0 to 329.999 mV
0 + 5 μV 20 μV
0.33 to 3.29999 V
0 + 20 μV 200 μV
3.3 to 7 V
0 + 100 μV 1000 μV
[1] Two channels of dc voltage output are provided.