AC Current, Square Wave Characteristics (typical)
Range LCOMP Risetime Settling Time Overshoot
I < 6 A @ 400 Hz
off
25 μs 40 μs to 1 % of final value < 10 % for < 1 V Compliance
3 A & 20 A Ranges on
100 μs 200 μs to 1 % of final value < 10 % for < 1 V Compliance
AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz Aberrations
0.3 % of p-p value, from 10 % to 90 % point
< 1 % of p-p value, with amplitude > 50 % of range