Manual Supplement 5520A Operator
23 11/10
Change #2
On page 4-7, following the first sentence, add the following note:
Note
If the 5500A is operated outside the range of t
cal
±
5 ºC, then the temperature coefficient defined in the
General Specifications, Chapter 1 of this manual, must be calculated and added to the absolute
uncertainties. Zeroing the 5500A is still required
Change #3, 55384, 55896, 55897, 55403, 55401
On page 8-6, Table 8-1, under Amplitude Characteristics, replace Range, with the following:
Range
0 to ±6.599 V 0 to ±130 V ±1 mV to ±6.599 V p-p ±1 mV to ±130 V p-p
On page 8-7, Table 8-2, replace Amplitude Range (p-p) and Frequency Range with the following:
Amplitude Range (p-p) 4.5 mV to 2.75 V
±(2 % of output + 200 μV)
Frequency Range
[1]
900 Hz 10 11 MHz
±(2.5 ppm of setting)
On page 8-8, Table 8-3, under Frequency Characteristics, replace Resolution with the following:
50 kHz (reference) 50 kHz to 100 MHz 100 MHz to 300 MHz 300 MHz to 600 MHz
Resolution 1 kHz 10 kHz
On page 8-10, Table 8-6, under Pulse Generator Characteristics replace Typical rise/fall times
and under Pulse Period replace Range with the following:
Typical rise/fall times <2 ns
Range 22 ms to 200 ns (45.5 Hz to 5 MHz)
On page 8-11, replace Table’s 8-7 and 8-9, with the following:
Pulse Period
Division Ratio
Amplitude into 50 Ω (p-p)
Typical Rise Time
22 ms to 200 ns off/1/10/100
1 V 2 ns
Edge Signal
Frequency
Division Ratio
Typical Amplitude into
50 Ω (p-p)
Typical Rise Time Typical Lead Time
900 Hz to 11 MHz off/1
1 V 2 ns
40 ns