5560A/5550A/5540A
Service Manual
8
Limits section in this manual.
Table 1. Required Equipment for Main Output
Equipment Description Manufacture/Model/PN Application
AC Measurement Standard Fluke Calibration 5790B Vac, Iac
Resistance Standards
Fluke 742A Series (100, 1 k, 10 k,
100 k, 1 M, 10 M)
Ohms, Idc
DC Reference Standard Fluke 732C Vdc
Frequency Counter Tektronix FCA3100 (Option MS or HS) Frequency
Shunts
Fluke A40B Series (1 mA, 10 mA,
100 mA, 1 A, 5 A, 10 A, 20 A, 50 A)
Idc, Iac, Phase
Calibrator (precision source) Fluke Calibration 5730A
Ohms, Temperature, Cap
Charge
Resistance Standard MI 9331G/100 M Ohms
Resistance Standard MI 9331G/1 G Ohms
Digital Multimeter Fluke Calibration 8588A
Ohms, Idc, Iac, Vdc, Vac,
Cap Charge
LCR Meter Hioki IM3533 Capacitance, Inductance
Phase meter Clarke-Hess 6000A Phase
Thermometer Fluke 1504 Temperature
Thermistor probe Fluke 5610-9-B Temperature
Type J thermocouple Omega TJ36-ICIN-18U-8-SMPW-M Temperature
Dewar flask with cap Various Temperature
Type B copper mini-plug
with copper wire
Various Temperature
BANANA/DBL BANANA, 2
COND 24 in
Pomona 1368-A-24 Idc, Vac
Low Thermal Cables
5730A-7003,TEST LEAD SET, LOW
THERMAL SPADE (Fluke Calibration
PN 4376018)
Various
Double banana RG58 cable Pomona Model 2BC-24 and 2BC-36 Ohms, Idc, Iac, Vdc, Vac