Calibration Verification of the Product
41
Amps AC Calibration
Table 15 lists the required equipment.
The amps ac calibration uses two methods. For frequencies
10 Hz, use a low-frequency Fluke
Calibration 8588A meter to monitor the output of Fluke A40B series shunts. For frequencies >10 Hz the
calibration is accomplished with the Fluke Calibration 5790B AC Measurement Standard monitoring
the output of the Fluke A40B series shunts.
Step
Product Output
(OUTPUT VZ)
Product Output
(AUX Output)
Product
Frequency
Product Output Limits
9 1 V 5 V 45 Hz ±4.05 mV
10 1 V 5 V 400 Hz ±4.05 mV
11 1 V 5 V 1 kHz ±4.05 mV
12 1 V 5 V 5 kHz ±7.9 mV
13 1 V 5 V 10 kHz ±15.4 mV
14 1 V 5 V 30 kHz ±188 mV
Table 15. Required Equipment for Amps AC
Equipment Description Manufacture/Model/PN Quantity
Digital Multimeter Fluke 8588A 1
(1 mA to 1.2 mA) Fluke A40B-1mA Current Shunt 1
(10 mA to 12 mA) Fluke A40B-10mA Current Shunt 1
(100 mA to 120 mA) Fluke A40B-100mA Current Shunt 1
(1 A to 1.2 A) Fluke A40B-1A Current Shunt 1
3.1 A, 3.11 A Fluke A40B-5A Current Shunt 1
10 A Fluke A40B-10A Current Shunt 1
(12.1 A to 20 A) Fluke A40B-20A Current Shunt 1
30 A Fluke A40B-50A Current Shunt 1
Type N (M) to dual banana (M) E-Z-Hook PN 9415 1
Type N (F) to Type N (F) adapter Various 1
LC Male to LC Male adapter A40B-ADAPT/LC 1
LC Female to N Male inter-series adapter A40B-ADAPT/LCN 1
N to 4 mm double banana connector A40B-LEAD/4mm 1
DBL BANANA N (M) RG58C/U ( 16 in) (Fluke PN 900394) 5790-8026 1