5560A/5550A/5540A
Service Manual
64
Frequency Calibration
Table 29 lists the required equipment.
Table 28. Phase Calibration Record for V/I
Step Product Output
Product
Frequency
Product
Phase
Shunt
Phase
Error
Measured
Product
Output
Calculated
Product
Output
Limits
8 10 V/0.1 A 65 Hz 0° ±0.10 °
9 10 V/0.1 A 65 Hz 30° ±0.10 °
10 10 V/0.1 A 65 Hz 45° ±0.10 °
11 10 V/0.1 A 65 Hz 60° ±0.10 °
12 10 V/0.1 A 65 Hz 90° ±0.10 °
13 10 V/0.1 A 65 Hz 180° ±0.10 °
14 10 V/0.1 A 65 Hz 270° ±0.10 °
15 10 V/0.1 A 400 Hz 90° ±0.25 °
16 10 V/0.1 A 1000 Hz 90° ±0.50 °
17 10 V/0.1 A 5000 Hz 90° ±2.5 °
18 10 V/0.1 A 10000 Hz 90° ±5.0 °
19 10 V/0.1 A 30000 Hz 90° ±10 °
20 10 V/3.11 A 65 Hz 0° ±0.10 °
21 10 V/3.11 A 400 Hz 0° ±0.25 °
22 1 V/12.1 A 65 Hz 180° ±0.10 °
23 10 V/12.1 A 65 Hz 30° ±0.10 °
24 1.21 V/20 A 65 Hz 0° ±0.10 °
25 10 V/20 A 400 Hz 90° ±0.25 °
26 250 V/0.1 A 65 Hz 90° ±0.10 °
27 250 V/0.1 A 10000 Hz 0° ±5.00 °
28 250 V/0.1 A 30000 Hz 45° ±10.00 °
Table 29. Required Equipment for Frequency Calibration
Equipment Description Manufacture/Model/PN Quantity
Frequency Counter Tektronix FCA3100 (Option MS or HS) 1
BNC (F) To Single Banana Plug Pomona Model 1894 1
BNC (M) To BNC (M) cable Various 1