Calibrator
Calibration Verification of the Product
17
Amps DC Calibration
Table 6 lists the required equipment.
To calibrate the Amps dc function (subsequent amps dc calibration steps refer to Table 7):
1. Set the 8588A as follows: Volts dc, Range Auto, NPLC 50, Guard ON (external guard).
2. Make sure that the DUT is in Standby.
3. Connect a strap between the Guard and Ground terminals of the DUT.
4. Complete an internal DC Zero Calibration on the Product.
5. Calculate the shunt resistance in
Ω from the calibration records and/or regression analysis for all
shunts listed in Table 7.
6. Record the shunt resistance in the
Ω column marked as Shunt_R in Table 7.
7. Connect the 8588A for V dc measurement with the low thermal cable to Fluke 742A-1k SENSE See
Figure 5.
Table 6. Required Equipment for Amps DC
Equipment Description Manufacture/Model/PN Quantity
Digital Multimeter Fluke Calibration 8588A 1
±(0 µA to 100 µA) Fluke 742A-1k Resistance Standard 1
±(1 mA to 1.2 mA) Fluke A40B-1mA Current Shunt 1
±(10 mA to 12 mA) Fluke A40B-10mA Current Shunt 1
±(100 mA to 120 mA) Fluke A40B-100mA Current Shunt 1
±(1 A to 1.2 A) Fluke A40B-1A Current Shunt 1
±3 A Fluke A40B-5A Current Shunt 1
±10 A Fluke A40B-10A Current Shunt 1
±(12 A to 20 A) Fluke A40B-20A Current Shunt 1
±30 A Fluke A40B-50A Current Shunt 1
Type N (M) to dual banana (M) E-Z-Hook PN 9415 1
Type N (F) to Type N (F) adapter Various 1
LC Male to LC Male adapter A40B-ADAPT/LC 1
LC Female to N Male inter-series adapter A40B-ADAPT/LCN 1
N to 4 mm double banana connector A40B-LEAD/4mm 1
DBL BANANA N (M) RG58C/U ( 16 in) (Fluke PN 900394) 5790-8026 1
Test Lead Banana to Banana 24 in Pomona 1368-A-24 1
Low Thermal Cables Fluke 5730-7003 1