5560A/5550A/5540A
Service Manual
24
Figure 14. Amps DC with A40B-50A
44. For all A dc steps, calculate the output with the formula below. Use the appropriate conversion to
convert from base units to units for the specific evaluation step.
45. With the entire table populated, evaluate each step for compliance to limits. If Guardbanding is to
be applied, create a table from the example in this manual, where the acceptance limits are
properly modified by your Guardbanding method.
Table 7. Calibration Steps for Amps DC
Step
Product
Range
Product
Output
Shunt
Shunt_R
[Ω]
DMM
Reading
on Shunt
Output
[V]
Calculated
Product
Output
Limits
1 120 µA 0 µA Fluke 742A-1k ±6.0 nA
2 120 µA -0.0001 µA Fluke 742A-1k ±6.0 nA
3 120 µA 10 µA Fluke 742A-1k ±7.0 nA
4 120 µA -10 µA Fluke 742A-1k ±7.0 nA
5 120 µA 100 µA Fluke 742A-1k ±16.0 nA
6 120 µA -100 µA Fluke 742A-1k ±16.0 nA
7 1.2 mA 0 mA Fluke 742A-1k ±15.0 nA
8 1.2 mA -0.000001 mA Fluke 742A-1k ±15.0 nA
5560A
8588A
INPUT
A40B-50A
OUTPUT
Calculated Product Output = DMM Reading on Shunt output [V]
Shunt_R [:]