Calibrator
Calibration Adjustment
67
6. Measure with the meter and input the value on the touch screen in the Enter Measured Value field.
Note that the Product only accepts values in base units.
7. Continue to follow the on-screen instructions OPERATE/Measure/Input Value until all steps in the
function are complete.
8. When a section is complete, the Product highlights the following section.
Volts AC Calibration Adjustment (OUTPUT VZ)
The Product begins the adjustment with internal steps. The volts ac calibration uses two methods. For
frequencies
10 Hz, a low frequency DMM (Fluke Calibration 8588A) is used. For frequencies >10 Hz,
use the Fluke Calibration 5790B AC Measurement Standard.
1. Set the 8588A as follows: Volts ac, Range Auto, Bandwidth Wide, Coupling: DC 10 M
Ω, Guard ON
(external guard)
2. Set the 5790B for Input 2, Auto range, High resolution ON, EXT GUARD.
3. When the Product displays a connect prompt, tap Continue.
4. At this time, the Product displays the amplitude/frequency.
a. For steps with frequency
10 Hz, connect the 8588A for v ac measurement as per Figure 17.
b. For steps with frequency
>10 Hz connect the 5790B for Vac measurement as per Figure 18.
5. Push OPERATE.
6. Measure with the meter and input the value on the touch screen in the Enter Measured Value field.
Note that the Product only accepts values in base units.
7. Continue to follow the on-screen instructions (OPERATE/Measure/Input Value) until all steps in the
function are complete. Alternate meters based on frequency of the step as for 4 a and b above.
8. When a section is complete, the Product highlights the following section.