Calibrator
Calibration Verification of the Product
53
Capacitance
Table 17 lists the required equipment.
Step
Product
Range
Product
Output
Product
Frequency
Shunt
Shunt_R
[Ω]
DMM
Reading on
Shunt
Output [V]
Calculated
Product
Output
Limits
78 12 A 10 A 1 kHz Fluke A40B-10A ±2.90 mA
79 12 A 10 A 5 kHz Fluke A40B-10A ±3.80 mA
80 12 A 10 A 10 kHz Fluke A40B-10A ±21.0 mA
81 30.2 A 12.1 A 3 Hz Fluke A40B-20A ±20 mA
82 30.2 A 12.1 A 45 Hz Fluke A40B-20A ±15 mA
83 30.2 A 12.1 A 1 kHz Fluke A40B-20A ±15 mA
84 30.2 A 12.1 A 5 kHz Fluke A40B-20A ±56 mA
85 30.2 A 20 A 3 Hz Fluke A40B-20A ±26 mA
86 30.2 A 20 A 45 Hz Fluke A40B-20A ±19 mA
87 30.2 A 20 A 1 kHz Fluke A40B-20A ±19 mA
88 30.2 A 20 A 5 kHz Fluke A40B-20A ±88 mA
89 30.2 A 30 A 3 Hz Fluke A40B-50A ±34 mA
90 30.2 A 30 A 45 Hz Fluke A40B-50A ±25 mA
91 30.2 A 30 A 1 kHz Fluke A40B-50A ±25 mA
92 30.2 A 30 A 5 kHz Fluke A40B-50A ±128 mA
Table 17. Required Equipment for Capacitance Calibration
Equipment Description Manufacture/Model/PN Quantity
Digital Multimeter Fluke 8588A 1
LCR Meter Hioki IM3533 1
Calibrator (precision current source) Fluke 5730A 1
BNC (M) to BNC (M) 4W lead set for LCR Various 1
Low Thermal Cables 5730A-7003 2
BNC (F) to BNC (M) Y Adapter Pomona Model 6700 2
BNC (F) To Single Banana Plug Pomona Model 1894 2
Table 16. Calibration Steps for Amps AC (cont.)