Calibration Verification of the Product
39
Volts AC Calibration (AUX Output) – 5550A and 5560A
Table 13 lists the required equipment.
To proceed with volts ac out of the Sense VZ LO/Output VI AUX proceed as follows:
1. Connect the Product to Fluke 5790B as shown in Figure 19.
2. Connect a strap between the guard and ground terminals of the DUT.
Step Product Range Product Output Product Frequency
Calculated
Product
Output
Limits
112 330 V 121 V 50 kHz ±37 mV
113 330 V 121 V 100 kHz ±158 mV
34 330 V 330 V 3 Hz ±735 mV
35 330 V 330 V 5 Hz ±306 mV
36 330 V 330 V 10 Hz ±46 mV
115 330 V 330 V 1 kHz ±46 mV
116 330 V 330 V 20 kHz ±46 mV
117 330 V 330 V 50 kHz ±87 mV
118 330 V 330 V 100 kHz ±409 mV
37 1020 V 331 V 3 Hz ±737 mV
38 1020 V 331 V 5 Hz ±307 mV
39 1020 V 331 V 10 Hz ±118 mV
120 1020 V 331 V 1 kHz ±118 mV
121 1020 V 331 V 10 kHz ±118 mV
122 1020 V 1000 V 40 Hz ±195 mV
123 1020 V 1000 V 1 kHz ±195 mV
124 1020 V 1020 V 10 kHz ±197 mV
Table 13. Required Equipment Volts AC from the AUX Output
Equipment Description Manufacture/Model/PN Quantity
AC Measurement Standard Fluke 5790B 1
Test Lead Banana to Banana 24 in Pomona Model 1368-A-24 1
Single banana to single banana lead Various 1
Table 12. Calibration Steps for Volts AC (cont.)