5560A/5550A/5540A
Service Manual
38
Step Product Range Product Output Product Frequency
Calculated
Product
Output
Limits
23 12 V 10 V 5 Hz ±7.75 mV
24 12 V 10 V 10 Hz ±1.50 mV
90 12 V 10 V 1 kHz ±1.20 mV
91 12 V 10 V 20 kHz ±1.20 mV
92 12 V 10 V 50 kHz ±2.45 mV
93 12 V 10 V 100 kHz ±5.73 mV
94 12 V 10 V 300 kHz ±16.6 mV
95 12 V 10 V 500 kHz ±16.6 mV
25 120 V 12.1 V 3 Hz ±31.7 mV
26 120 V 12.1 V 5 Hz ±16.0 mV
27 120 V 12.1 V 10 Hz ±4.89 mV
97 120 V 12.1 V 1 kHz ±1.89 mV
98 120 V 12.1 V 20 kHz ±1.89 mV
99 120 V 12.1 V 50 kHz ±3.40 mV
100 120 V 12.1 V 100 kHz ±8.03 mV
28 120 V 100 V 3 Hz ±208 mV
29 120 V 100 V 5 Hz ±78 mV
30 120 V 100 V 10 Hz ±15.00 mV
102 120 V 100 V 40.01 Hz ±12.00 mV
103 120 V 100 V 1 kHz ±12.00 mV
104 120 V 100 V 20 kHz ±12.00 mV
105 120 V 100 V 50 kHz ±25 mV
106 120 V 100 V 100 kHz ±57 mV
31 330 V 121 V 3 Hz ±317 mV
32 330 V 121 V 5 Hz ±160 mV
33 330 V 121 V 10 Hz ±22 mV
108 330 V 121 V 1 kHz ±22 mV
109 330 V 121 V 10 kHz ±22 mV
110 330 V 121 V 10.01 kHz ±22 mV
111 330 V 121 V 20 kHz ±22 mV
Table 12. Calibration Steps for Volts AC (cont.)