Calibrator
Calibration Verification of the Product
37
Step Product Range Product Output Product Frequency
Calculated
Product
Output
Limits
68 1.2 V 0.121 V 1 kHz ±74 µV
69 1.2 V 0.121 V 20 kHz ±74 µV
70 1.2 V 0.121 V 50 kHz ±43 µV
71 1.2 V 0.121 V 100 kHz ±108 µV
72 1.2 V 0.121 V 300 kHz ±264 µV
73 1.2 V 0.121 V 500 kHz ±264 µV
16 1.2 V 1 V 3 Hz ±2.08 mV
17 1.2 V 1 V 5 Hz ±0.77 mV
18 1.2 V 1 V 10 Hz ±0.18 mV
75 1.2 V 1 V 40.01 Hz ±0.12 mV
76 1.2 V 1 V 1 kHz ±0.12 mV
77 1.2 V 1 V 20 kHz ±0.12 mV
78 1.2 V 1 V 50 kHz ±0.25 mV
79 1.2 V 1 V 100 kHz ±0.60 mV
80 1.2 V 1 V 300 kHz ±1.60 mV
81 1.2 V 1 V 500 kHz ±1.60 mV
19 12 V 1.21 V 3 Hz ±3.17 mV
20 12 V 1.21 V 5 Hz ±1.60 mV
21 12 V 1.21 V 10 Hz ±0.49 mV
83 12 V 1.21 V 1 kHz ±0.19 mV
84 12 V 1.21 V 20 kHz ±0.19 mV
85 12 V 1.21 V 50 kHz ±0.34 mV
86 12 V 1.21 V 100 kHz ±0.80 mV
87 12 V 1.21 V 300 kHz ±2.54 mV
88 12 V 1.21 V 500 kHz ±2.54 mV
22 12 V 10 V 3 Hz ±20.75 mV
Table 12. Calibration Steps for Volts AC (cont.)