Calibration Verification of the Product
33
Volts AC Calibration (OUTPUT VZ)
Table 11 lists the required equipment.
Step
Product
Range
Product
Output
DMM Mode/
Method
Transfer
Resistor
Transfer
Value
DMM
Measure-
ment on
Transfer
Resistor
DMM
Measure-
ment on
Product
Calculated
Product
Output
Limits
22
1.2 G
Ω 0.121 GΩ
Normal 100
M
Ω / Transfer
MI
9331G/
100 M
M
Ω MΩ MΩ
GΩ
±0.57 M
Ω
23
1.2 GΩ 0.3 GΩ
Normal 1 GΩ /
Transfer
N/A
Table 9,
Meter
error
µ
Ω/Ω GΩ GΩ ±1.27 MΩ
24
1.2 G
Ω 0.4 GΩ
Normal 1 GΩ /
Transfer
N/A
Table 9,
Meter
error
µ
Ω/Ω GΩ GΩ ±1.66 MΩ
25
1.2 G
Ω 0.64 GΩ
Normal 1 GΩ /
Transfer
N/A
Table 9,
Meter
error
µ
Ω/Ω GΩ GΩ ±2.6 MΩ
26
1.2 G
Ω 1 GΩ
Normal 1 GΩ /
Transfer
MI
9331G/
1 G
Table 9,
Meter
error
µΩ/Ω GΩ GΩ ±4.0 MΩ
27
12
Ω 0 Ω
Ext. current
source 100 µA
N/A N/A N/A
±40.0 m
Ω
28
120
Ω 100 Ω
Ext. current
source 100 µA
N/A N/A N/A
±42.2 m
Ω
Table 11. Required Equipment Volts AC Main Output
Equipment Description Manufacture/Model/PN Quantity
AC Measurement Standard Fluke 5790B 1
Digital Multimeter Fluke 8588A 1
Double banana RG58 cable Pomona Model 2BC-24 1
Single banana to single banana lead Various 1
Table 10. Calibration Steps for Resistance (cont.)