5560A/5550A/5540A
Service Manual
52
Step
Product
Range
Product
Output
Product
Frequency
Shunt
Shunt_R
[Ω]
DMM
Reading on
Shunt
Output [V]
Calculated
Product
Output
Limits
44 1.2 A 0.121 A 3 Hz
Fluke A40B-
100mA
±0.12 mA
50 1.2 A 0.121 A 45 Hz
Fluke A40B-
100mA
±0.07 mA
51 1.2 A 0.121 A 1 kHz
Fluke A40B-
100mA
±0.07 mA
52 1.2 A 0.121 A 5 kHz
Fluke A40B-
100mA
±0.10 mA
53 1.2 A 0.121 A 10 kHz
Fluke A40B-
100mA
±0.54 mA
54 1.2 A 0.121 A 30 kHz
Fluke A40B-
100mA
±0.78 mA
55 1.2 A 1 A 3 Hz Fluke A40B-1A ±0.30 mA
57 1.2 A 1 A 45 Hz Fluke A40B-1A ±0.25 mA
58 1.2 A 1 A 1 kHz Fluke A40B-1A ±0.25 mA
59 1.2 A 1 A 5 kHz Fluke A40B-1A ±0.28 mA
60 1.2 A 1 A 10 kHz Fluke A40B-1A ±2.30 mA
61 1.2 A 1 A 30 kHz Fluke A40B-1A ±4.30 mA
56 3.1 A 1.21 A 3 Hz Fluke A40B-1A ±0.86 mA
62 3.1 A 1.21 A 45 Hz Fluke A40B-1A ±0.59 mA
63 3.1 A 1.21 A 1 kHz Fluke A40B-1A ±0.59 mA
64 3.1 A 1.21 A 5 kHz Fluke A40B-1A ±0.66 mA
65 3.1 A 1.21 A 10 kHz Fluke A40B-1A ±2.92 mA
66 3.1 A 3.1 A 3 Hz Fluke A40B-5A ±1.43 mA
67 3.1 A 3.1 A 45 Hz Fluke A40B-5A ±1.04 mA
68 3.1 A 3.1 A 1 kHz Fluke A40B-5A ±1.04 mA
69 3.1 A 3.1 A 5 kHz Fluke A40B-5A ±1.23 mA
70 3.1 A 3.1 A 10 kHz Fluke A40B-5A ±6.70 mA
71 12 A 3.11 A 3 Hz Fluke A40B-5A ±1.93 mA
72 12 A 3.11 A 45 Hz Fluke A40B-5A ±1.25 mA
73 12 A 3.11 A 1 kHz Fluke A40B-5A ±1.25 mA
74 12 A 3.11 A 5 kHz Fluke A40B-5A ±1.73 mA
75 12 A 3.11 A 10 kHz Fluke A40B-5A ±7.22 mA
76 12 A 10 A 3 Hz Fluke A40B-10A ±4.00 mA
77 12 A 10 A 45 Hz Fluke A40B-10A ±2.90 mA
Table 16. Calibration Steps for Amps AC (cont.)