Calibrator
Calibration Verification of the Product
51
Step
Product
Range
Product
Output
Product
Frequency
Shunt
Shunt_R
[Ω]
DMM
Reading on
Shunt
Output [V]
Calculated
Product
Output
Limits
34 12 mA 10 mA 1 kHz
Fluke A40B-
10mA
±3.00 μA
35 12 mA 10 mA 5 kHz
Fluke A40B-
10mA
±3.00 μA
36 12 mA 10 mA 10 kHz
Fluke A40B-
10mA
±13.00 μA
37 12 mA 10 mA 30 kHz
Fluke A40B-
10mA
±50 μA
32 120 mA 12.1 mA 3 Hz
Fluke A40B-
10mA
±12 μA
38 120 mA 12.1 mA 45 Hz
Fluke A40B-
10mA
±6 μA
39 120 mA 12.1 mA 1 kHz
Fluke A40B-
10mA
±6 μA
40 120 mA 12.1 mA 5 kHz
Fluke A40B-
10mA
±10 μA
41 120 mA 12.1 mA 10 kHz
Fluke A40B-
10mA
±25 μA
42 120 mA 12.1 mA 30 kHz
Fluke A40B-
10mA
±148 μA
43 120 mA 100 mA 3 Hz
Fluke A40B-
100mA
±30 μA
45 120 mA 100 mA 45 Hz
Fluke A40B-
100mA
±17 μA
46 120 mA 100 mA 1 kHz
Fluke A40B-
100mA
±17 μA
47 120 mA 100 mA 5 kHz
Fluke A40B-
100mA
±28 μA
48 120 mA 100 mA 10 kHz
Fluke A40B-
100mA
±130 μA
49 120 mA 100 mA 30 kHz
Fluke A40B-
100mA
±500 μA
Table 16. Calibration Steps for Amps AC (cont.)