EasyManua.ls Logo

Fluke 5560A - Page 54

Fluke 5560A
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5560A/5550A/5540A
Service Manual
50
Step
Product
Range
Product
Output
Product
Frequency
Shunt
Shunt_R
[Ω]
DMM
Reading on
Shunt
Output [V]
Calculated
Product
Output
Limits
19 1.2 mA 0.121 mA 10 kHz
Fluke A40B-
1mA
±245 nA
20 1.2 mA 0.121 mA 30 kHz
Fluke A40B-
1mA
±5484 nA
4 1.2 mA 1 mA 3 Hz
Fluke A40B-
1mA
±300 nA
21 1.2 mA 1 mA 45 Hz
Fluke A40B-
1mA
±300 nA
22 1.2 mA 1 mA 1 kHz
Fluke A40B-
1mA
±300 nA
23 1.2 mA 1 mA 5 kHz
Fluke A40B-
1mA
±300 nA
24 1.2 mA 1 mA 10 kHz
Fluke A40B-
1mA
±1300 nA
25 1.2 mA 1 mA 30 kHz
Fluke A40B-
1mA
±9.00 μA
5 12 mA 1.21 mA 3 Hz
Fluke A40B-
1mA
±1.24 μA
26 12 mA 1.21 mA 45 Hz
Fluke A40B-
1mA
±1.24 μA
27 12 mA 1.21 mA 1 kHz
Fluke A40B-
1mA
±1.24 μA
28 12 mA 1.21 mA 5 kHz
Fluke A40B-
1mA
±1.24 μA
29 12 mA 1.21 mA 10 kHz
Fluke A40B-
1mA
±2.45 μA
30 12 mA 1.21 mA 30 kHz
Fluke A40B-
1mA
±14.84 μA
31 12 mA 10 mA 3 Hz
Fluke A40B-
10mA
±3.00 μA
33 12 mA 10 mA 45 Hz
Fluke A40B-
10mA
±3.00 μA
Table 16. Calibration Steps for Amps AC (cont.)

Table of Contents