Calibrator
Calibration Verification of the Product
49
Table 16. Calibration Steps for Amps AC
Step
Product
Range
Product
Output
Product
Frequency
Shunt
Shunt_R
[Ω]
DMM
Reading on
Shunt
Output [V]
Calculated
Product
Output
Limits
1 120 μA10 μA3 Hz
Fluke A40B-
1mA
±12.0 nA
6 120 μA10 μA45 Hz
Fluke A40B-
1mA
±12.0 nA
7 120 μA10 μA 1 kHz
Fluke A40B-
1mA
±12.0 nA
8 120 μA10 μA 5 kHz
Fluke A40B-
1mA
±12.0 nA
9 120 μA10 μA10 kHz
Fluke A40B-
1mA
±52.0 nA
10 120 μA10 μA30 kHz
Fluke A40B-
1mA
±1040.0 nA
2 120 μA100 μA3 Hz
Fluke A40B-
1mA
±30.0 nA
11 120 μA100 μA45 Hz
Fluke A40B-
1mA
±30.0 nA
12 120 μA100 μA 1 kHz
Fluke A40B-
1mA
±30.0 nA
13 120 μA100 μA 5 kHz
Fluke A40B-
1mA
±30.0 nA
14 120
μA100 μA10 kHz
Fluke A40B-
1mA
±160.0 nA
15 120 μA100 μA30 kHz
Fluke A40B-
1mA
±1400 nA
3 1.2 mA 0.121 mA 3 Hz
Fluke A40B-
1mA
±124 nA
16 1.2 mA 0.121 mA 45 Hz
Fluke A40B-
1mA
±124 nA
17 1.2 mA 0.121 mA 1 kHz
Fluke A40B-
1mA
±124 nA
18 1.2 mA 0.121 mA 5 kHz
Fluke A40B-
1mA
±124 nA