5560A/5550A/5540A
Service Manual
18
Figure 5. Amps DC with 742A-1k - Meter Zero
8. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
9. Connect the test equipment as shown in Figure 6.
10. Output from the DUT/Measure with the meter and record the 8588A measured values in the DMM
Reading on Shunt Output [V] column. Complete all steps for the present connection (1 to 20). Refer
to the Product manual how to set and use range lock on bottom of range outputs.
11. Set the Product to Standby.
12. Move the red lead connected to the Product V AUX to the 30 A terminal.
13. Complete the subsequent steps for this connection (21 and 22, Table 7). See the Product manual
for how to set and use range lock on bottom of range outputs.
Figure 6. Amps DC with742A-1k
742A-1k
Resistance Standard
CURRENT SENSE
HI HI
LO LO
NOMINAL VALUE = 1 K:
TEMPERATURE RANGE: 18 °C TO 28 °C
MAXIMUM VOLTAGE = 10 VDC
MAXIMUM CURRENT = 0.01 A
CHASSIS
GROUND
742A-1k
8588A
742A-1k
Resistance Standard
CURRENT SENSE
HI HI
LO LO
NOMINAL VALUE = 1 K:
TEMPERATURE RANGE: 18 °C TO 28 °C
MAXIMUM VOLTAGE = 10 VDC
MAXIMUM CURRENT = 0.01 A
CHASSIS
GROUND
5560A
8588A
742A-1k