Calibrator
Calibration Adjustment
71
Table 31. Calibration Adjust Steps for Capacitance with LCR meter
Step Product Output Test Frequency
1 0.2 nF 1000 Hz
2 1 nF 1000 Hz
3 1.21 nF 1000 Hz
4 10 nF 1000 Hz
5 12.1 nF 610 Hz
6 20 nF 610 Hz
7 30 nF 610 Hz
8 50 nF 610 Hz
9 100 nF 610 Hz
10 120 nF 610 Hz
11 0.121 µF 100 Hz
12 1 µF 100 Hz
13 1.21 µF 80 Hz
14 10 µF 80 Hz
15 12.1 µF 20 Hz
16 100 µF 20 Hz
17 0.121 mF 5 Hz
18 1 mF 5 Hz
19 1.21 mF 2 Hz
20 10 mF 2 Hz
21 12.1 mF 1 Hz
22 100 mF 1 Hz