5730A
Calibration Manual
36
15. To reject the changes, touch Cancel. Otherwise, touch Continue to accept
and save the changes, and to continue with calibration.
Once the internal steps are complete, a screen similar to this is shown:
Setup Menu Calibration
Artifact Calibration
Calibration Complete
The largest change was 30% of spec.
22VDC
Would you like to save changes?
DiscardSave
hhp215.eps
16. Touch Save to store the results and adjust the Product. If the security
passcode has not already been entered, the Product prompts to do so before
the results are saved. Touch Discard to remove the results from the
calibration.