5730A
Calibration Manual
46
The shifts of references and outputs are given both as absolute shifts in V, A,
ohms, and also (for non-zero magnitudes) as shifts relative to the previous value,
in parts per million (ppm) or percent (%).
The specified accuracy for the output is also shown and the ratio of the shift to
that accuracy in percent.
The Calibration Shift report file name is CSyymmdd.CSV, where yymmdd is the
date today.
Calibration Check Shift Results
The Calibration Check report presents changes in the magnitude of self-
correction generated by calibration check. Calibration check uses internal
standards rather than external standards as the basis for comparison and does
not make permanent changes to calibration constants. This report is available
only after a calibration check has been done and until power has been reset to
the Product.
The report contains this information:
Heading that shows the date and time the report was saved and user report
string
Modules present (installed or attached)
Most recent calibration dates and temperatures
Values of externally-calibrated internal references
DC voltage output shifts for each range
AC voltage output shifts for each range
AC and dc current output shifts for each range
Resistance shifts for each value
5725A Boost Amplifier output shifts (if attached).
The calibration report file name is CKyymmdd.CSV, where yymmdd is the date
today.
Raw Data Results
This report gives the values for the internally stored correction factors (calibration
constants), and is useful primarily for Fluke Calibration service centers.